YC-TI-PS-0006 | Ceyear 87246

YC-TI-PS-0006 | Ceyear 87246 Series LAN Peak/Average Power Sensor

Price on request

Request a quote

The 87246 series provides average and pulse power measurement and CCDF analysis over LAN, with PoE power and communication. Models D/F/L have different frequency limits and power ranges, supporting remote measurement, external triggering and buffered acquisition.

LAN power sensor series appearance reference showing multiple models, not an included set. The supplied model and connector are defined in the quotation.

Models and measurement ranges

Manufacturer modelFrequency rangeAverage power measurement accuracy
87246D50 MHz–18 GHz±0.20 dB
87246F50 MHz–40 GHz±0.30 dB
87246L500 MHz–67 GHz±0.33 dB

Power ranges and modes

Manufacturer modelMeasurement modeFrequency intervalPower measurement range
87246D / 87246FNormal mode50 MHz≤f<500 MHz−30…+20 dBm
87246D / 87246FNormal modef≥500 MHz−35…+20 dBm
87246D / 87246FAverage modeFull selected-model band−45…+20 dBm
87246LNormal mode500 MHz≤f≤40 GHz−35…+20 dBm
87246LNormal mode40 GHz<f≤67 GHz−30…+20 dBm
87246LAverage mode500 MHz≤f≤40 GHz−45…+20 dBm
87246LAverage mode40 GHz<f≤67 GHz−40…+20 dBm

Accuracy conditions

Accuracy applies to average power only: −15…+20 dBm, f≥500 MHz, DUT SWR<1.20, free-running mode and averaging value 32. It is not a guarantee for every peak or pulse measurement.

Maximum VSWR

Manufacturer modelFrequency intervalMaximum VSWR
87246D / 87246F50 MHz≤f≤2 GHz1.20
87246D / 87246F2 GHz<f≤18 GHz1.26
87246F18 GHz<f≤26.5 GHz1.35
87246F26.5 GHz<f≤40 GHz1.50
87246L500 MHz≤f≤2 GHz1.20
87246L2 GHz<f≤18 GHz1.26
87246L18 GHz<f≤26.5 GHz1.35
87246L26.5 GHz<f≤40 GHz1.50
87246L40 GHz<f≤67 GHz1.70

Connection and core functions

Power and communicationCompatible PoE connection required
Power meter compatibilityNot compatible with the 2439D microwave power meter
Remote measurement distanceUp to 60 m with a compatible PoE connection and suitable network cable
Video bandwidth≥30 MHz
Rise/fall time≤13 ns
Maximum continuous sample rate80 MSamples/s
Minimum pulse width50 ns
Horizontal scale2 ns/div–100 ms/div
Internal trigger range−20…+20 dBm (typical)
  • D/F start at 50 MHz and L at 500 MHz; each power interval is also limited by the selected model upper frequency.
  • Sixteen pulse power and time parameters are measured. Up to 50,000 readings/s applies to buffered mode, not every trigger and transfer setting.
  • Horizontal scale is 2 ns/div–100 ms/div; the USB 87236 limit of 3600 s/div does not apply.
  • An ordinary Ethernet port does not replace the required PoE connection. For multiple sensors, check power budget and trigger arrangement; no unspecified synchronization accuracy is assumed.

Options and quotation

PoE switch 87240-H01 and 1.5/3/5/15/30/60 m shielded Category 6 cables are options; the switch is not included by default. Specify model-specific -H12 English options, report and extended-warranty requirements when requesting a quote. Provide model, signal type, power range, network environment, accessories and quantity. Scope of supply and delivery schedule are defined in the quotation.

Request a quote