YC-TI-MUL-0002 | Ceyear 82401/06 Series Millimeter-Wave Frequency Multiplier Module
Price on request
Request a quoteThe 82401/06 series multiplies the output of compatible signal generators into model-specific bands across 50–1100 GHz. It uses standard waveguide outputs and a 15 V power adapter, with fixed saturated-power output for millimeter-wave and terahertz signal generation.
82401/06 series appearance reference. Port layout may vary with band module; the supplied model and waveguide are defined by the configuration table.
Models and frequency bands
Output power specifications are measured at 25±5 °C and apply to the corresponding model.
| Manufacturer model | Output frequency | Input frequency | Multiplier | Output waveguide | Saturated output power |
|---|---|---|---|---|---|
| 82406 | 50–75 GHz | 12.5–18.8 GHz | 4 | WR15 | ≥+13 dBm |
| 82401N | 60–90 GHz | 10.0–15.0 GHz | 6 | WR12 | ≥+11 dBm |
| 82406A | 75–110 GHz | 12.5–18.4 GHz | 6 | WR10 | ≥+10 dBm |
| 82401QA | 90–140 GHz | 10.0–15.6 GHz | 9 | WR8.0 | ≥+5 dBm |
| 82406B | 110–170 GHz | 9.1–14.2 GHz | 12 | WR6.5 | ≥+2 dBm |
| 82406C | 170–220 GHz | 14.1–18.4 GHz | 12 | WR5.1 | ≥−2 dBm |
| 82401SA | 170–260 GHz | 9.4–14.5 GHz | 18 | WR4.3 | ≥−6 dBm |
| 82406D | 220–325 GHz | 12.2–18.1 GHz | 18 | WR3.4 | ≥−8 dBm |
| 82401TA | 260–400 GHz | 9.6–14.9 GHz | 27 | WR2.8 | ≥−12 dBm |
| 82406E | 325–500 GHz | 9.0–13.9 GHz | 36 | WR2.2 | ≥−18 dBm |
| 82401U | 500–750 GHz | 10.4–15.7 GHz | 48 | WR1.5 | ≥−25 dBm |
| 82401V | 750–1100 GHz | 10.4–15.3 GHz | 72 | WR1.0 | ≥−33 dBm |
Technical specifications
| Output mode | Fixed saturated-power output; no levelled-power output |
|---|---|
| Drive signal power | +5…+10 dBm; frequency within the selected model input range |
| RF drive input connector | 3.5 mm female |
| Power supply | 15 V power adapter |
| Power consumption | ≤30 W |
| Weight | ≤5 kg |
Usage and ordering
- 50–1100 GHz is the combined coverage of twelve models, not continuous full-range coverage by one module.
- Levelled-power output is not supported. Programmable levelling capabilities of the 1400 series do not apply to this series.
- Saturated output power specifications are measured at 25±5 °C. Use the input frequency range and output waveguide specified for the selected model.
- 82406-JC, 82401N-JC and 82406A-JC are manufacturer test-service options with a test report; they are not automatically included in every quotation.
Quotation and delivery
Specify the output band, compatible signal generator model, required output power, waveguide interface and quantity. Band modules, accessories, test services and delivery schedule are defined in the quotation.