YC-TI-RCV-0001 | 3916H EMI test receiver, series appearance example

YC-TI-RCV-0001 | 3916 Series EMI Test Receiver

Price on request

Request a quote

The 3916 series EMI test receiver supports EMC pre-compliance, testing against relevant standards and interference analysis, with stepped scanning, FFT time-domain scanning and APD measurement. Five frequency configurations offer port 1 upper limits of 8.4, 18, 26.5, 45 or 50 GHz. See the model table for frequency ranges and connectors. Real-time spectrum and wideband IQ recording depend on selected options.

Model selection

ModelPort 1 DC coupledPort 1 AC coupledPort 1 RF connector
3916B1 Hz–8.4 GHz10 MHz–8.4 GHzN female,50 Ω
3916D1 Hz–18 GHz10 MHz–18 GHzN female,50 Ω
3916E1 Hz–26.5 GHz10 MHz–26.5 GHz3.5 mm male,50 Ω
3916F1 Hz–45 GHz10 MHz–45 GHz2.4 mm male,50 Ω
3916H1 Hz–50 GHz10 MHz–50 GHz2.4 mm male,50 Ω

Common specifications

ParameterSpecification
Port 2 frequency rangeDC: 1 Hz–1 GHz; AC: 9 kHz–1 GHz
Port 2 connectorN female, 50 Ω; surge-protected reception channel
10 MHz precision reference accuracy±(time since last calibration × aging rate + temperature stability + calibration accuracy)
Reference aging±3×10⁻⁸/year
Reference temperature stability±1×10⁻⁹
Reference calibration accuracy±5×10⁻⁹
Frequency resolution0.1 Hz
Preselection channels23 bands, including notch filter
Frequency readout accuracy±(frequency reading × reference accuracy + 0.1% span + 5% RBW + 2 Hz + 0.5 × horizontal resolution); horizontal resolution = span/(sweep points − 1)
Marker count accuracy±(marker frequency × reference accuracy + 0.100 Hz)
Counter resolution0.001 Hz
Dwell time50 μs–100 s
Sweep timeSpan ≥10 Hz: 3 μs–16000 s; zero span: 1 μs–16000 s; zero-span accuracy ±0.01%
Maximum scan-list segments20
DetectorsMax peak, min peak, RMS, quasi-peak, average, CISPR average and CISPR RMS
Phase noise1 GHz carrier, 20–30 °C: <-107 dBc/Hz at 100 Hz offset; <-125 at 1 kHz; <-133 at 10 kHz; <-135 at 100 kHz
Residual FM≤0.10 Hz × N; RBW and VBW 10 Hz, nominal over 20 ms; N depends on frequency
Residual FM multiplier N2 Hz<f≤8.2 GHz: 1; 8.2<f≤18 GHz: 1; 18<f≤26.5 GHz: 2; 26.5<f≤40 GHz: 2; 40<f≤50 GHz: 4
Time-domain FFT bandwidth29.8 MHz standard; 390 MHz optional
EMI filter bandwidth at −6 dB1 Hz, 10 Hz, 100 Hz, 200 Hz, 1 kHz, 9 kHz, 10 kHz, 100 kHz, 120 kHz, 1 MHz, 10 MHz; accuracy ≤3% except 10 MHz, ≤5% at 10 MHz; selectivity −60/−6 dB ≤4.5:1
Standard filter bandwidth at −3 dB1 Hz–3 MHz in 1/2/3/5 steps, plus 4/5/6/8/10/20 MHz; accuracy at 3.01 dB: ±3% for 1 Hz≤RBW≤1 MHz, ±15% for 1 MHz<RBW≤20 MHz; selectivity −60/−3 dB <5.0:1
Video bandwidth1 Hz–20 MHz, 1/2/3/5 steps; nominal
1 dB compression pointTwo-tone test, RF gain prioritized for dynamic range, tone separation >5 × prefilter bandwidth; preselector/preselection amplifier/LNA off: ≥8 dBm at 20 MHz≤f<3.25 GHz; ≥7 dBm at 3.25 GHz≤f≤50 GHz
Third-order interceptRF gain prioritized for dynamic range, preselector/preselection amplifier/LNA off: ≥20 dBm at 10 MHz≤f≤100 MHz; ≥18 at 100 MHz<f≤3.25 GHz; ≥20 at 3.25<f≤8.2 GHz; ≥18 at 8.2<f≤50 GHz
Second-harmonic interceptPreselector/preselection amplifier/LNA off: ≥45 dBm at 10 MHz≤f≤4.1 GHz; ≥60 at 4.1<f≤13.25 GHz; ≥55 at 13.25<f≤20 GHz; ≥45 at 20<f≤25 GHz
Residual response0 dB input attenuation: ≤-90 dBm at f≤1 MHz; ≤-103 dBm at 1 MHz<f≤8.4 GHz
Absolute amplitude uncertainty≤0.2 dB at internal 500 MHz calibration signal, 20–30 °C
Image rejection≥80 dBc
Maximum CW input level+30 dBm only with preselector and preamplifier off and input attenuation ≥20 dB
Input VSWRPort 1, DC coupling, attenuation ≥10 dB: ≤1.2 at 9 kHz≤f≤1 GHz; ≤1.5 at 1<f≤8 GHz; ≤1.8 at 8<f≤45 GHz; ≤2.0 at 45<f≤50 GHz
Input attenuator0–70 dB, 2 dB steps
Communication interfacesUSB-A: 3 front and 3 rear; RJ45 LAN 10/100/1000M auto-negotiation; GPIB IEEE-488, 24-pin connector
Audio/video interfaces3.5 mm headphone; DP video
Timebase interfaces1–100 MHz reference input: BNC female, 50 Ω; 10 MHz output: BNC female, 50 Ω; 1 GHz input: SMA female, 50 Ω; 100 MHz/1 GHz output: SMA female, 50 Ω
Trigger interfacesTrigger inputs 1/2 and outputs 1/2: BNC female
Dimensions W × H × D including handles and feet(474.6±4) × (237.5±2.5) × (559.5±4) mm
Dimensions W × H × D excluding handles and feet(426±4) × (221.5±2.5) × (450±4) mm
Weight<35 kg
Power supply100–240 VAC, 50/60 Hz, automatic
Maximum power consumption600 W
Operating temperature0–50 °C
Storage temperature-40–70 °C

Functions and measurements

ParameterSpecification
APD measurement bandwidths200 Hz, 9 kHz, 120 kHz, 1 MHz
APD measurement rangeMinimum measurable probability 10⁻⁷; maximum measurement time 120 s
FFT processingUp to 16384 points; 50–13000 detector channels per FFT step; overlap >93.75%, amplitude error within 0.4 dB
Real-time spectrum40/200/400 MHz by option; signal duration for 100% POI better than 0.31 μs, subject to configuration
EMI diagnosticsOver-limit peak search, limit margin and multi-detector final measurements from a list
Spectrum measurementsChannel power, occupied bandwidth and adjacent-channel power
Typical performanceDANL down to -163 dBm with amplifier on; TOI +27 dBm; phase noise -135 dBc/Hz at 1 GHz carrier and 10 kHz offset. Typical values do not replace the condition-dependent specification tables
Preselection amplifierApprox. 20 dB gain (design value), distinct from the optional H34 low-noise preamp with approx. 30 dB design gain
Noise and frequency response

A–H identify measurement configurations, not additional product models.

ConditionH34 low-noise preamp optionPreselectorPreselection amplifierLNA
ANot fittedOffOffNot applicable
BNot fittedOnOffNot applicable
CNot fittedOnOnNot applicable
DFittedOffOffOff
EFittedOnOffOff
FFittedOnOnOff
GFittedOffOffOn
HFittedOnOffOn

Displayed average noise level (DANL): input terminated in 50 Ω, 0 dB input attenuation, +20 to +30 °C, normalized to 1 Hz. Values below are maximum levels in dBm and apply only within the selected model frequency range. A–H refer to the configuration conditions shown below.

Frequency rangeABCDEFGH
10 MHz≤f<100 MHz≤-148≤-150≤-151≤-145≤-149≤-150≤-155≤-156
100 MHz≤f<1.2 GHz≤-147≤-153≤-155≤-146≤-152≤-154≤-160≤-157
1.2 GHz≤f<2.2 GHz≤-146≤-153≤-155≤-145≤-152≤-154≤-159≤-157
2.2 GHz≤f<3.25 GHz≤-145≤-152≤-154≤-144≤-151≤-153≤-159≤-157
3.25 GHz≤f<5.25 GHz≤-143≤-152≤-152≤-142≤-151≤-151≤-157≤-156
5.25 GHz≤f<6.5 GHz≤-144≤-153≤-153≤-142≤-152≤-152≤-156≤-157
6.5 GHz≤f<8.2 GHz≤-138≤-151≤-151≤-137≤-150≤-150≤-156≤-156
8.2 GHz≤f<18 GHz≤-141≤-141≤-141≤-140≤-140≤-140≤-156≤-155
18 GHz≤f<26.5 GHz≤-141≤-141≤-141≤-140≤-140≤-140≤-154≤-155
26.5 GHz≤f<40 GHz≤-137≤-137≤-137≤-136≤-136≤-136≤-153≤-152
40 GHz≤f≤50 GHz≤-132≤-132≤-132≤-131≤-131≤-131≤-148≤-148

Frequency response: 10 dB input attenuation, 20–30 °C, DC coupled, relative to the 500 MHz calibration frequency. Values are ±dB. The 1 Hz–9 kHz entries for A–F are nominal; other entries retain the stated conditions. G/H cover only frequencies from 10 MHz.

Frequency rangeABCDEF
1 Hz≤f<9 kHz±1 (nominal)±1.3 (nominal)±1.7 (nominal)±1.1 (nominal)±1.4 (nominal)±1.8 (nominal)
9 kHz≤f<10 MHz±0.6±0.9±1.3±0.7±1±1.4
10 MHz≤f<3.6 GHz±0.5±0.8±1.2±0.6±0.9±1.3
3.6 GHz≤f<8 GHz±0.6±1±1.6±0.7±1.1±1.7
8 GHz≤f<26.5 GHz±1.5±1.5±2±1.6±1.6±1.6
26.5 GHz≤f≤50 GHz±2±2±2.8±2.1±2.1±2.1
Frequency rangeGH
10 MHz≤f<3.6 GHz±1±1.3
3.6 GHz≤f<8 GHz±1.3±1.7
8 GHz≤f<26.5 GHz±2±2
26.5 GHz≤f≤50 GHz±2.8±2.8
Optional functions and services
Option codeFunction or serviceConfiguration
3916-H02Auxiliary IF outputSecond IF, 1 Hz resolution, 15 dB gain in 1 dB steps; 425 MHz±40 MHz for analysis bandwidth ≤40 MHz, 750 MHz±600 MHz for 200–400 MHz
3916-H12CWideband digital interfaceOptical-fibre IQ output up to 400 MHz; for 4712C recorder
3916-H12EWideband digital interfaceOptical-fibre IQ output up to 1.2 GHz; for 4712E recorder
3916-H13Precision frequency referenceHigher-accuracy frequency reference
3916-H22C-4T4712C data recorder4 TB; requires H12C; IQ recording up to 400 MHz
3916-H22C-8T4712C data recorder8 TB; requires H12C; IQ recording up to 400 MHz
3916-H22C-16T4712C data recorder16 TB; requires H12C; IQ recording up to 400 MHz
3916-H22C-32T4712C data recorder32 TB; requires H12C; IQ recording up to 400 MHz
3916-H22E-8T4712E data recorder8 TB; IQ recording up to 1.2 GHz; match receiver digital interface and recorder configuration
3916-H22E-16T4712E data recorder16 TB; IQ recording up to 1.2 GHz; match receiver digital interface and recorder configuration
3916-H22E-32T4712E data recorder32 TB; IQ recording up to 1.2 GHz; match receiver digital interface and recorder configuration
3916-H22E-64T4712E data recorder64 TB; IQ recording up to 1.2 GHz; match receiver digital interface and recorder configuration
3916-H33-8Electronic attenuator9 kHz–8 GHz; 30 dB range, 0.5 dB steps
3916-H34-08Low-noise preamplifierApprox. 30 dB gain (design value); select for receiver upper frequency
3916-H34-18Low-noise preamplifierApprox. 30 dB gain (design value); select for receiver upper frequency
3916-H34-26Low-noise preamplifierApprox. 30 dB gain (design value); select for receiver upper frequency
3916-H34-45Low-noise preamplifierApprox. 30 dB gain (design value); select for receiver upper frequency
3916-H34-50Low-noise preamplifierApprox. 30 dB gain (design value); select for receiver upper frequency
3916-H40External frequency extensionExternal mixing, LO output/IF input and signal identification; not available for 3916B; range depends on separately purchased module
3916-H41-4040 MHz real-time spectrumDigital persistence, seamless waterfall and frequency-mask trigger
3916-H41-200200 MHz real-time spectrumDigital persistence, seamless waterfall and frequency-mask trigger
3916-H41-400400 MHz real-time spectrumDigital persistence, seamless waterfall and frequency-mask trigger
3916-H52A390 MHz time-domain FFT scanFFT fast scanning up to 390 MHz
3916-H53A-40390 MHz FFT plus 40 MHz real-time spectrumFFT fast scanning up to 390 MHz and digital persistence at the selected bandwidth
3916-H53A-200390 MHz FFT plus 200 MHz real-time spectrumFFT fast scanning up to 390 MHz and digital persistence at the selected bandwidth
3916-H53A-400390 MHz FFT plus 400 MHz real-time spectrumFFT fast scanning up to 390 MHz and digital persistence at the selected bandwidth
3916-H96Printed user manualDetailed printed manual
3916-H97Rack-mount kitRack handles and accessories
3916-H98English kitEnglish panel, manual, user interface and operating system
3916-H99Aluminium transport caseHandle and wheels
3916-H99-2Plastic wheeled safety caseHandle and wheels
3916-S04Phase-noise measurementSingle-sideband phase-noise curve and single-point measurement
3916-S09Analogue demodulationAM/FM/φM modulation and distortion analysis
3916B-EWT1One-year extended warrantyOne year beyond standard warranty; two items for two years. Excludes calibration; one-way freight only. Availability and terms subject to quotation
3916D-EWT1One-year extended warrantyOne year beyond standard warranty; two items for two years. Excludes calibration; one-way freight only. Availability and terms subject to quotation
3916E-EWT1One-year extended warrantyOne year beyond standard warranty; two items for two years. Excludes calibration; one-way freight only. Availability and terms subject to quotation
3916F-EWT1One-year extended warrantyOne year beyond standard warranty; two items for two years. Excludes calibration; one-way freight only. Availability and terms subject to quotation
3916H-EWT1One-year extended warrantyOne year beyond standard warranty; two items for two years. Excludes calibration; one-way freight only. Availability and terms subject to quotation
3916B-JLMetrology serviceThird-party calibration and certificate; inclusion subject to order
3916D-JLMetrology serviceThird-party calibration and certificate; inclusion subject to order
3916E-JLMetrology serviceThird-party calibration and certificate; inclusion subject to order
3916F-JLMetrology serviceThird-party calibration and certificate; inclusion subject to order
3916H-JLMetrology serviceThird-party calibration and certificate; inclusion subject to order

Standard accessories

ParameterSpecification
Power cordStandard three-core cord
Quick-start guideQuick-start documentation
Product certificateFactory product conformity document

Selection and use

  • Provide required upper frequency, applicable test standard, signal level, detector and recording bandwidth to select receiver and options. Applications include relevant GB/T6113, GJB151 and CISPR-series testing.
  • Standard time-domain FFT bandwidth is 29.8 MHz. 390 MHz FFT, real-time spectrum, wideband IQ interfaces and recorders require matched options and are not included by default.
  • The +30 dBm maximum CW input level applies only to the specified front-end state. Check attenuation, amplifiers and input protection before connecting a signal.
  • Frequency and performance tables apply only within the selected receiver frequency range. Nominal, design and typical values are not guaranteed specifications for every configuration.
  • External extension option H40 is unavailable for 3916B; extension modules are purchased separately. Match recorders, digital interfaces and low-noise preamps to receiver frequency and bandwidth.
  • The photo shows a 3916H appearance example. Supplied model, options, accessories, calibration and extended-warranty services are defined by quotation and ordered configuration.

Request receiver selection and quotation